5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

512

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

5.6.3.2.4

DET Ipol

5.6.3.2.4.1

DET Ipol - Overview

Dialog window: Method

▶ DET Ipol ▶ Properties... ▶ DET Ipol - 'Command name'

Command for Dynamic equivalence point titrations with voltametric
measurement (selectable polarization current).

Devices
This command can be executed with the following instruments:

Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the DET Ipol command are set on the following
seven tabs:

■

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

■

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

■

Titration parameters
Parameters for the run of the titration.

■

Stop conditions
Definition of conditions which cause the titration to stop.

■

Potentiometric evaluation
Parameters for the potentiometric evaluation of titration curves.