6 tet - thermometric evaluation, 7 tet - additional evaluations – Metrohm tiamo 2.4 (process analysis) User Manual

Page 941

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

925

5.6.3.9.6

TET - Thermometric evaluation

Tab: Method

TET Properties... Thermometric evaluation

Command name

Name of the command.

Entry

25 characters

Parameters for the thermometric evaluation of the titration curve with
automatic endpoint recognition.

Evaluation start

Possibility to postpone the start of the evaluation. The evaluation starts
not before having added the solution volume.

Input range

0.0 - 50.0 mL

Default value

0.1 mL

Endpoints
Selection of the sorting for the display of the endpoints in the table.

Selection

Sort by volume (ascending) | Sort by peak
height (descending)

Default value

Sort by volume (ascending)

Table of endpoints

A maximum of 9 endpoints can be defined. The parameters for determin-
ing these endpoints are displayed in the window table which cannot be
edited and which contains the columns Reaction type and EP criterion.

[New]

Open the dialog window Endpoint # in which the parameters for a new
endpoint can be entered (see Chapter 5.6.3.9.9, page 928).

[Properties]

Open the dialog window Endpoint # in which the parameters for the
selected endpoint can be edited (see Chapter 5.6.3.9.9, page 928).

[Delete]

Delete the endpoint selected in the table.

5.6.3.9.7

TET - Additional evaluations

Tab: Method

TET Properties... Additional evaluations

Command name

Name of the command.

Entry

25 characters