5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1059

Command variables
The following command variables are generated by the MEAS Opt Conc
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':

Identifica-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.DBL

Total duration for the processing of the command in s

.EME

End measured value (measured value after processing of
the command) in the unit of the measured value

.ETE

1)

End temperature (temperature after the command has
been processed) in °C

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command
has never ended

.IME

Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value

.ITE

1)

Initial temperature (temperature before start conditions
are processed) in °C

.LP.CAx

Calculated value x (1 - 3) for the last measuring point on
the measuring point list

.LP.EXx

External value x (1 - 3) for the last measuring point on
the measuring point list

.LP.CNT

Intensity of the last measuring point in the measuring
point list in counts

.LP.MEA

Measured value for the last measuring point in the mea-
suring point list in the unit of the measured value

.LP.TIM

Time in s until the last measuring point in the measuring
point list is reached

.LP.TRN

Transmission of the last measuring point in the measur-
ing point list in %

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off
)

.NMP

Number of measuring points in measuring point list