5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

783

5.6.3.4.6.10

SET Upol - Fixed endpoint evaluation

Dialog window: Method

▶ SET Upol ▶ Properties... ▶ Additional evalua-

tions

▶ Fixed endpoint evaluation ▶ [New/[Properties] ▶ Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-200.0 - 200.0 µA

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

5.6.3.5

KFT

5.6.3.5.1

KFT - Overview

Menu item: Method

▶ Insert ▶ New command...

Command for volumetric Karl Fischer titrations.

Principle
Volumetric titrations for determining the water content according to Karl
Fischer with automatic conditioning are carried out with this command. A
typical feature of the Karl Fischer titration is the controlled reagent dosing,
which depends on the control difference (current measured value – end-
point), with the aim of reaching the measured value given as being the
predefined target endpoint as quickly and accurately as possible. Overti-
trating is avoided to a great extent. The titration termination at the end-
point takes place either drift-controlled or after a waiting period. The vol-
ume added up to the endpoint gives the calculable reagent consumption.

Reagent dosing takes place in three phases during the titration:

■

Initial dosing
The dosing rate increases continuously during this phase. It starts with
the Min. rate and increases to the Max. rate.