7 Devices

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1627

Depending on the type the function and the graphical representation of
the port varies.

Selection

not defined | Pump | Stirrer | valve | Gen. out-
put
| Alarm signal

Default value

not defined

NOTE

Additionally a stepping motor can have a port of the Type = Brake.

The port information for this port type is automatically read out of the
stepping motor controller.

Default value

Display of the default value the outputs takes directly after starting and
exiting tiamo. The default value is also set after an error or when cancel-
ing the method.

NOTE

Behavior at Emergency stop:

All digital outputs except the ones of the Type = Alarm signal are
set to the default value when an emergency stop takes place.

A digital output of the Type = Alarm signal is set to the non-
default value
when an emergency stop takes place.

Digital outputs assigned to a digital input of the Type = Level
display
or Leak detector do not change their status when an
emergency stop takes place.

Selection

active | inactive

Default value

inactive

Default values for commands

Parameter 1

The meaning of Parameter 1 depends on the Type of the digital output
that has been set.

When Type = Stirrer: stirrer output in %, pulse duration dependent on
Parameter 2.

Input range

-214783647 - 214783647

Default value

100