Metrohm tiamo 2.4 (process analysis) User Manual
Page 1210
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1194
■■■■■■■■
tiamo 2.4 (for Process Analysis)
Titrino
Input range
0.01 - 150.00 mL/min
Default value
maximum mL/min
NOTE
The dosing rate should be reduced for viscous liquids.
Dosing time
Time during which dosing is to be carried out. This parameter is only dis-
played for Dosing criterion = Volume/Dosing time or Dosing rate/
Dosing time.
Input range
0 - 9999999 s
Default value
100 s
Pause
Waiting period before dosing is started.
Input range
0 - 999999 s
Default value
0 s
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously. This value is used for
temperature correction during monitoring with pH measurements..
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Input range
0.1 - 999999 s
Default value
5.0 s