5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
799
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando
Input range
0.1 - 999,999 s
Default value
2.0 s
Titrino
Input range
0.08 - 16,200 s
Default value
2.00 s
5.6.3.5.2.6
KFT Ipol - Stop conditions
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.
Stop volume
Stop when the entered volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample size or the titration vessel size.
Titrando
Input range
0.00000 - 9,999.99 mL
Default value
100.000 mL
Selection
off
Titrino
Input range
0.00 - 9,999.99 mL
Default value
100.00 mL
Selection
off