5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

799

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando
Input range

0.1 - 999,999 s

Default value

2.0 s

Titrino
Input range

0.08 - 16,200 s

Default value

2.00 s

5.6.3.5.2.6

KFT Ipol - Stop conditions

Tab: Method

KFT Ipol Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.

Stop volume

Stop when the entered volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample size or the titration vessel size.

Titrando
Input range

0.00000 - 9,999.99 mL

Default value

100.000 mL

Selection

off

Titrino
Input range

0.00 - 9,999.99 mL

Default value

100.00 mL

Selection

off