6.7 Sensors subwindow
■■■■■■■■■■■■■■■■■■■■■■
1446
■■■■■■■■
tiamo 2.4 (for Process Analysis)
Input range
–2.147E+9 - 2.147E+9 mV
Default value
–30 mV
Upper limit
Upper intervention limit for E(0).
Input range
–2.147E+9 - 2.147E+9 mV
Default value
30 mV
Warning limits for cell constant (for conductivity sensors
only)
on | off (Default value: off)
If this option is activated, then the values for the cell constant will be
shown in orange if the limits are exceeded.
Lower limit
Lower warning limit for the cell constant.
Input range
0.001 - 500 /cm
Default value
2.0 /cm
Upper limit
Upper warning limit for the cell constant.
Input range
0.001 - 500 /cm
Default value
400 /cm
Intervention limits for cell constant (for conductivity sensors
only)
on | off (Default value: off)
If this option is activated, then the values for the cell constant will be
shown in red if the limits are exceeded.
Lower limit
Lower intervention limit for the cell constant.
Input range
0.001 - 500 /cm
Default value
0.001 /cm
Upper limit
Upper intervention limit for the cell constant.
Input range
0.001 - 500 /cm
Default value
500 /cm