5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1109

Data transmission from standard
Input of the data of the conductivity standard, which should be as close as
possible to the Current temperature. This data is as a rule contained in
the Data Sheet/Certificate for the conductivity standard.

Reference temperature

Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.

Input range

-20.0 - 150.0°C

Default value

20.0

Conductivity

Input of the conductivity of the standard at the reference temperature.

Input range

0.00000 - 2000.00 mS/cm

Default value

1.0 mS/cm

Temperature coefficient

Temperature coefficient of the standard with the reference temperature.

Input range

0.00 - 9.99 %/°C

Default value

1.00 %/°C

5.6.5.5

CAL Spec

5.6.5.5.1

CAL Spec - Overview

Dialog window: Method

CAL Spec Properties... CAL Spec - 'Command

name'

Command for wavelength calibration.

Devices
This command can be executed with the following instrument:

Spectrometer: Avantes