6 Configuration

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1435

for ISE (+4)
Input range

–2.147E+9 - 2.147E+9 mV

Default value

17.0 mV

for ISE (–4)
Input range

–2.147E+9 - 2.147E+9 mV

Default value

–12.0 mV

pH(0) monitoring (for pH electrodes only)
on | off (Default value: off)

If this option is activated, then a check will be made to see whether the
electrode zero point pH(0) determined during the calibration of a pH elec-
trode is within the defined limit values. If it is outside them, a correspond-
ing message appears and you can select whether you will nevertheless
accept the value or whether the previous value is to be retained.

Lower limit

Lower limit value for pH(0).

Input range

–2.147E+9 - 2.147E+9

Default value

6.750

Upper limit

Upper limit value for pH(0).

Input range

–2.147E+9 - 2.147E+9

Default value

7.250

E(0) monitoring (for ISE electrodes only)
on | off (Default value: off)

If this option is activated, then a check will be made to see whether the
electrode zero point E(0) determined during the calibration of an ISE elec-
trode is within the defined limit values. If it is outside them, a correspond-
ing message appears and you can select whether you will nevertheless
accept the value or whether the previous value is to be retained.

Lower limit

Lower limit value for E(0).

Input range

–2.147E+9 - 2.147E+9 mV

Default value

–2,000.0 mV

Upper limit

Upper limit value for E(0).