5 set ph - titration parameters, Titration parameters – Metrohm tiamo 2.4 (process analysis) User Manual

Page 725

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

709

Titrino
Input range

-20.00 - 20.00

Selection

off

Default value

off

For additional parameters for Endpoint 2, see Endpoint 1.

5.6.3.4.3.5

SET pH - Titration parameters

Tab: Method

SET pH Properties... Titration parameters

Command name

Name of the command.

Entry

25 characters

Parameters for the run of the titration.

Titration direction

Selection of the titration direction. If two endpoints are set, then the titra-
tion direction is already established and the following selection of titra-
tion direction
is ignored.

Selection

+ | - | auto

Default value

auto

+
positive measured value alteration
-
negative measured value alteration
auto
The titration direction is determined automatically from the start mea-
sured value and the set endpoint.

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). For example, it makes sense to
enter an extraction time for the titration of samples that are not easily
soluble.

Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Input range

0 - 999999 s

Temperature

Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware
tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.

Advertising