6 Configuration

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1411

Input range

1 - 9

Default value

2

Dosing rate

Rate at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out, the rate will automatically be reduced to
the highest possible value.

Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

Filling rate

Rate at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the
function is carried out, the rate will automatically be reduced to the high-
est possible value.

Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

Maximum dosing / filling rate for the exchange unit, depending
on the cylinder volume:

Cylinder volume

Maximum rate

1 mL

3.0 mL/min

5 mL

15.0 mL/min

10 mL

30.0 mL/min

20 mL

60.0 mL/min

50 mL

150.0 mL/min

NOTE

Enter lower rates for solutions of higher viscosity.