5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

829

Command variables
The following command variables are generated in the method run by the
command KFC and can be used in formulas under the designation 'Com-
mand name.Variable designation'
:

Designation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CBY

Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated

.COK

Command status; 1 = Condition requirement ful-
filled
, 0 = Condition requirement not fulfilled

.DBL

Total duration for the processing of the command in s

.DRI

Current and/or last drift for drift correction in µL/min

.DTI

Time for drift correction (time from the start of the
titration to the end of the command) in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.EP.CHA

Charge for endpoint in mA·s

.EP.DVT

Drift for the endpoint in µg/min

.EP.MEA

Measured value for the endpoint in mV

.EP.QTY

Measured value (water) for the endpoint in µg

.EP.TEM

Temperature for the endpoint 1 in °C

.EP.TIM

Time in s until the endpoint is reached

.EP.TIM

Time in s until the endpoint is reached

.ETE

End temperature (temperature following processing of
the command) in °C

.FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) = Com-
mand has never ended

.FP{x}.CHA

Charge for the fixed endpoint × (1…9) in mA·s

.FP{x}.DVT

Drift for the fixed endpoint × (1…9) in µg/min

.FP{x}.MEA

Measured value for the fixed endpoint × (1…9) in the
unit of the measured value