5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1227

Pumps

Tower

Selection of the tower of the Sample Processor for moving to the required
lift position. Both Tower 1 and Tower 2 can always be selected even if the
Sample Processor has only one tower.

Selection

1 | 2

Default value

1

Pump(s)

Selection of the pump that is to be actuated. The pump(s) can either be
built-in or externally connected. With 1+2 both pumps at the selected
tower will be switched at the same time.

Selection

1 | 2 | 1+2

Default value

1

1
Pump 1 is actuated.
2
Pump 2 is actuated.
1+2
Both pumps at the selected tower will be switched at the same time.

Actions

Selection

Switch on | Switch off | Duration

Default value

Switch on

Switch on
Switch on the pump(s).
Switch off
Switch off the pump(s).
Duration
Switch on the pump(s) for a particular period.

Input range

0 - 9999.9

Default value

10.0

Selection

s | min

Default value

s