5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

976

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Window

A range (window) can be defined on the measured value axis or on the
time axis. The break point evaluation will only be carried out in the
defined window. Only the first break point in the defined window will be
recognized.

Selection

Measured value | Time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Time for the lower limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mV

Default value

-2000.0 mV

Window = Time
Input range

0 - 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Value for the upper limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mV

Default value

2000.0 mV

Window = Time
Input range

1 - 999999 s

Default value

999999 s

5.6.4.5.5

MEAS Ipol - Additional measured values

Tab: Method

MEAS Ipol Properties... Additional measured values

Command name

Name of the command.

Entry

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in
the default settings in additional measured value columns Calculated #
or External # in the measuring point list.