Metrohm tiamo 2.4 (process analysis) User Manual

Page 1710

Advertising
background image

8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1694

■■■■■■■■

tiamo 2.4 (for Process Analysis)

NOTE

Before a sample spectrum can be measured, a reference spectrum has
to be recorded in a determination by a spectrometer using a MEAS
Ref
command.

In the case of a one-channel instrument setup, the parameters
Start wavelength, End wavelength, Integration time, Averaged
spectra
, Smoothing and Flash rate are read out from the reference
spectrum and displayed but they cannot be modified. In the case of a
two-channel instrument setup, these parameters are editable,
except for Integration time and Smoothing.

Start wavelength

Lower limit of the spectrum.

Input range

100.0 - 2000.0 nm

Default value

400.0 nm

End wavelength

Upper limit of the spectrum.

Input range

100.0 - 2000.0 nm

Default value

1000.0 nm

Integration time

Integration time for spectrum recording.

Input range

0.01 - 600000 ms

Default value

6 ms

Averaged spectra

Number of spectra that are recorded and averaged.

Input range

1 - 10000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.

Input range

0 - 100 Pixels

Default value

0 Pixels

Flash rate

Number of flashes per second of a pulsed xenon lamp as light source.