Changing the Mixer Barrel

The mixer module is used to blend buffers from two to four separate solutions. It is

important to use a mixer with a barrel size that is based on the flow rate of

separation. Bio-Rad offers several different mixer barrel sizes for different flow rates.

The F10 systems ship with the 263

μ

l mixer base and top assembly. When using a

higher flow rate, the mixer volume of this base unit can be expanded by adding a

750

μl

or 2 ml barrel. The F100 systems ship with 750

μl

mixer base and top

assembly. The mixer volume can be increased at higher flow rates by inserting

either 2ml, 5 ml, or 12 ml barrel.

Table

10 lists the recommended barrel size for different flow rates for systems

without a buffer blending valve (gradient mode).

Table

11 on page

99 lists the recommended barrel size for different flow rates for

systems with a buffer blending valve.

Tip:

For liquids that are difficult to mix, use a mixer barrel with a higher

capacity.

Table 10.

Recommended NGC system mixer barrel sizes —

without buffer blending valve (gradient)

Barrel Size

Flow Rate, ml/min

F10 Pump

F100 Pump

263 μl

(Base and top)

0.1–1.00

n/a

750 μl

(Base and top)

1.00–9.99

1.00–4.99

2 ml

10.00–20.00

5.00–14.99

5 ml

n/a

15.00–39.99

12 ml

n/a

40.00–200.00