Bio-Rad ChromLab™ Software User Manual

Page 41

Advertising
background image

Valves

Instrument Guide

| 39

Sample Inject Valve LEDs

The sample inject valve module is fitted with LEDs.

Solid green LEDs indicate the flow is from the system pump.

Solid blue LEDs indicate the flow is from the sample pump.

Blinking green LEDs indicate the line to plumb using the Point-to-Plumb™

feature.

How the Sample Inject Valve Works

The sample inject valve enables sample to be loaded onto the column. Several

sample application techniques are available:

The sample loop can be manually filled with sample using a syringe

followed by injection onto a column by the system pumps

The sample loop can be filled with sample using the sample pump (if

installed) followed by injection onto a column by the system pumps

Sample can be injected directly onto the column using the sample pump (if

installed)

Flow Paths of the Sample Inject Valve

In Manual mode (from the System Control tab in ChromLab), you can manually

change the flow path of the sample inject valve from its dialog box.

The following images show the different flow paths of the sample inject valve.

L

EGEND

Manual injection flow

System pump flow

Sample pump flow

No flow

Advertising