Bio-Rad ChromLab™ Software User Manual

Page 64

Advertising
background image

2

| The NGC Instrument

62

| NGC Chromatography Systems and ChromLab Software

During operation, the system pressure sensor continuously measures and records

the pressure. If the system pressure exceeds user- or system-specified pressure

limits, the system pumps shut down. Alternatively, when you select the option in

ChromLab to control the flow to avoid overpressure, the NGC system reduces the

flow rate until the system is within safe operation pressure limits. See

F10 Pump on

page

30

and

F100 Pump on page

30

for specific pressure limitations.

L

EGEND

1

Mixer outlet port

2

Mixer barrel

3

Inlet from system pump B

4

Mixer base assembly

5

Inlet from system pump A

6

Pressure transducer port