Bio-Rad ChromLab™ Software User Manual
Page 179
Advertising
System and Sample Pumps
Instrument Guide
| 177
4.
Position the left wash housing with its arrow pointing up and carefully slide it
into place over the piston.
5.
Position the right wash housing with its arrow pointing down and carefully slide
it into place over the piston.
6.
Connect a pump wash tubing to the top left and right wash housing ports.
7.
Connect an inlet tube to the bottom left wash housing inlet port.
8.
Connect a priming tube to the bottom right wash housing port.
9.
Tighten the tubing fittings with a fittings tightener.