| Troubleshooting the Instrument

248

| NGC Chromatography Systems and ChromLab Software

The pumps do not
deliver the correct flow
rate (continued).

The pump seals or check valves
might be fouled.

Tip:

Replace the check

valves on a regular basis,
depending on the usage and
type of solution or sample
pumped through the pump
head. See Table

14

, NGC

instrument maintenance
schedule, on page

152

.

Use an ultrasonic cleaner and
a mild detergent solution and
then rinse with water to
recondition the pump seals
and check valves.

Perform CIP to clean the
pump seals and check
valves.

See the NGC Chromatography
Systems and ChromLab
Software User Guide for more
information.

Replace the pump seals.

See

Disassembling the Pumps

on page

159

.

Replace the check valves.

See

Disassembling the Pumps

on page

159

.

Note:

Recalibrate the

pumps after cleaning the
system or replacing the
check valves. In
ChromLab, select Tools >
Calibrate > Pump Flow
Rate. Complete the
calibration steps in the
dialog box.

Liquid continuously
leaks from the pump
head washout drain
trough.

The pump head seals are worn or
damaged.

Replace the pump head seals
on both pump heads.

Note:

Recalibrate the

pumps after changing the
seals. In ChromLab, select
Tools > Calibrate > Pump
Flow Rate. Complete the
calibration steps in the
dialog box.

Symptom

Possible Cause

Possible Solution