| Maintaining the Instrument

150

| NGC Chromatography Systems and ChromLab Software

Cleaning the Outer Surfaces of the NGC Instrument

During normal operation, spills and splashes can cause residue to form on the

module faceplates and fluidic lines.

Use a damp cloth to clean the faceplates. Avoid using excessive moisture around

the power switch and connector outlets located on the lower right and back sides of

the unit.

Clean the buffer tray. Rinse away any residue with water and empty through the

drainage hole.

Cleaning the NGC Fluidics System

Tip:

You can use the System CIP/Storage phase in ChromLab™ software to

clean the system’s fluidic path.

See

the

NGC Chromatography Systems and

ChromLab Software User Guide

for information about performing CIP (clean in

place)

.

Wash all fluidic lines with filtered 1 N NaOH solution.

Rinse the fluidic lines with filtered deionized water to remove all traces of NaOH.

Immerse all inlet lines in fresh water and rinse their external surface to remove

any trace of NaOH.