System and Sample Pumps

Instrument Guide

| 161

5.

Locate the wash housing O-ring and the pump head small washer. If you do not
see them on the separated pump head:

The O-ring might be attached to the wash housing.

The small washer might have remained on the pump piston.

6.

Remove and store the O-ring and small washer in a safe place such as a petri
dish or other small container.

Tip:

You can remove the O-ring from its groove with a pair of tweezers if

necessary.

Note:

Do not lose or damage the O-ring or small washer. The pump will

not operate without them.

7.

(Optional) Remove the inlet and outlet check valves on the pump head.

Tip:

It is not necessary to remove the inlet and outlet check valves unless

you plan to clean or replace them. See

To remove the pump head check

valves on page

162

.

8.

(Optional) Replace the pump head seal. For more information, see

To replace

the pump head seal on page

167

.