Detectors

Instrument Guide

| 71

The Backpressure Regulator

A backpressure regulator can be added to the flow path to generate backpressure

and prevent air bubbles from forming in the UV flow cells when medium pressure

columns are used. (The NGC systems ship with a 40 psi backpressure regulator.)

Bio-Rad recommends that the backpressure regulator be plumbed between the

conductivity monitor and the pH monitor to avoid exceeding the 70 psi limit of the

pH probe.

Note:

Bio-Rad does not recommend using backpressure regulators for

columns with a precolumn pressure limit less than 72 psi (0.5 MPa).