B troubleshooting the instrument, System and sample pumps, Appendix b – Bio-Rad ChromLab™ Software User Manual

Page 247: Troubleshooting the instrument, Btroubleshooting the instrument

Advertising
background image

Instrument Guide

| 245

B

Troubleshooting the
Instrument

This appendix lists potential problems and suggested solutions for the NGC™

chromatography systems. You can obtain more information about your system by

registering for Consult Bio-Rad at www.bio-rad.com, where an online technical

support service offers an extensive database of frequently asked questions (FAQs).

Note:

To troubleshoot other Bio-Rad instruments and devices, such as the

BioFrac™ fraction collector and the C-96 autosampler, refer to their separate

documentation.

System and Sample Pumps

Symptom

Possible Cause

Possible Solution

The pumps do not
deliver the correct flow
rate.

The pumps might be out of
calibration.

Recalibrate the pumps. Start
ChromLab™ software and
select Tools > Calibrate Pump
Flow Rate. Complete the
calibration steps in the dialog
box.

The incorrect pump heads might
be mounted.

In ChromLab, verify that the
fluidic scheme shows the
correct configuration.

Verify that you are using the
correct pump heads for the
method.

Advertising