Using the C-96 Autosampler with NGC Systems

Instrument Guide

| 293

To prime and purge the C-96 autosampler

Note:

Before priming the autosampler, ensure that any column is offline.

1.

Immerse the inlet tubing in a container of buffer or HPLC grade (filtered,
degassed) or other high quality water.

2.

Use a syringe to prime the pumps. See the section

Priming and Purging the

Systems on page

118

for specific information.

3.

On the ChromLab™ computer or the touch screen, start ChromLab in Manual
mode.

4.

Set the fluidic scheme to display the autosampler and sample inject valve.

5.

In the fluidic scheme, touch or click the Sample Inject Valve module to open its
dialog box.

6.

Set the inject valve to System Pump Inject Loop.

Autosampler