Other Components

Instrument Guide

| 237

To remove the backpressure regulator

1.

Disconnect the tubing from the top and bottom of the installed backpressure
regulator.

2.

Store the backpressure regulator in a safe place.

3.

Do one of the following:

Attach a Tefzel union to the open tubing in place of the backpressure

regulator.

Replace the two open tubes with a single tubing line.

Attach one end of the tubing to the top of the conductivity monitor and the

other end to either the pH flow cell or the BioFrac™ fraction collector.