| The NGC Instrument

32

| NGC Chromatography Systems and ChromLab Software

Piston Wash System

The pumps on the NGC system automatically rinse behind the piston with a

constant low flow of 20% ethanol. This prolongs the working lifetime of the seal by

keeping the seal wetted and preventing salt crystals from depositing onto the

pistons.

When the pumps are running, the pistons and the check valve located in each

chamber automatically pump the rinse fluid, which circulates through the system.

The tube holders on the side of the instrument hold two conical rinsing solution

reservoirs, one for the system pumps and one for the sample pump (if present). The

inlet and outlet tubing for the system pumps piston wash housing is immersed in

one of the rinsing solution reservoirs. The inlet and outlet tubing for the sample

pump is immersed in the second rinsing solution reservoir. The tubing connects to

the wash housing ports at the back of the pump heads.

Important:

Change the rinse fluid weekly. Place the washing system reservoir

at the same height or above the level of the pump heads to avoid siphoning of

the solution back to the reservoir. Ensure that the inlet tubing reaches the very

bottom of the rinsing solution reservoir.