Uv or uv/vis detector – Bio-Rad ChromLab™ Software User Manual

Page 255

Advertising
background image

UV or UV/Vis Detector

Instrument Guide

| 253

UV or UV/Vis Detector

Symptom

Possible Cause

Possible Solution

The UV baseline is
unstable or noisy.

The UV baseline shows
a reproducible zig-zag
or sawtooth trace.

Air bubbles are trapped in the flow
cell.

Note:

Always degas buffers

and solutions before use.

Degas buffers by stirring
vigorously under vacuum for
approximately 20 min.

Important:

Use a

heavy-wall, side-arm
Erlenmeyer flask, as
standard flasks may
implode under vacuum.

Plumb the detector with the
column outlet connected to
the bottom of the flow cell.
This forces air bubbles to rise
to the top of the flow cell and
dissipate.

Use the 40 psi backpressure
regulator supplied with the
NGC system to remove most
air bubbles. Place it directly
after the conductivity flow cell
and run buffer for several
minutes.

If the chosen column cannot be
run with the backpressure
regulator, and air bubbles are a
continuing problem (for example,
during some affinity
chromatography steps), use the
2 mm path length flow cell. The
straight line flow of the flow cell
makes it easier to remove
bubbles.

Advertising