Replacing the Conductivity Monitor

The images in this section show the conductivity monitor on the single-wavelength

UV detector module. Though the position of the conductivity monitor differs on the

multi-wavelength UV/Vis module, the steps to replace the conductivity monitor are

the same.

To replace the conductivity monitor

1.

On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.

2.

Disconnect the tubing lines from the conductivity monitor.

3.

Loosen the two thumbscrews on the front of the detector.