Eplace them. see, To replace the wash housing check valves – Bio-Rad ChromLab™ Software User Manual

Page 175

Advertising
background image

System and Sample Pumps

Instrument Guide

| 173

To replace the wash housing check valves

1.

Identify the left and right wash housings.

The wash system has two wash housings for each pump. The left housing

has check valves connected at the inlet and outlet ports. It is installed on

the pump with the arrow (flow direction) pointing up.

The right wash housing has empty fittings connected at the inlet and outlet

ports. The right housing is installed with the arrow (flow direction) pointing

down.

Advertising