How they work – Bio-Rad ChromLab™ Software User Manual

Page 72

Advertising
background image

2

| The NGC Instrument

70

| NGC Chromatography Systems and ChromLab Software

How They Work

UV Detectors

ChromLab recognizes the connected optic module (that is, either the

single-wavelength UV or multi-wavelength UV/Vis module). The lamps turn on when

the NGC instrument powers up. At the start of the method run, the system

determines if the monitor is ready.

The warmup time for the LED lamp in the single-wavelength detector is negligible.

The tungsten and deuterium lamps in the multi-wavelength detector require 60–90

min to warm up.

Note:

For most accurate data with the tungsten and deuterium lamps, allow a

minimum of 60 min warmup time.

Conductivity Monitor

The temperature-compensated conductivity monitor calculates the conductivity of

the sample and buffer mixture flowing through its flow cell. The stronger the salt

concentration, the higher the conductivity signal.

Since conductivity of a solution can vary with changes in temperature, temperature

variations of the sample are measured by the temperature sensor located in the

conductivity flow cell. A temperature compensation factor (set in ChromLab) can be

used to compare the conductivity of the sample in relation to the corresponding

conductance at room temperature or a user-defined reference temperature.

Advertising