Flow paths of the column switching valve – Bio-Rad ChromLab™ Software User Manual

Page 61

Advertising
background image

Valves

Instrument Guide

| 59

The section

Flow Paths of the Column Switching Valve

explains the specific flow

paths.

The column switching valve has integrated precolumn and postcolumn pressure

monitors. The precolumn pressure monitor measures the pressure that the system

exerts on the column (precolumn pressure). The postcolumn pressure monitor

measures the pressure at the column outlet and provides the delta pressure across

the column. In Manual and Method modes, you can set alarms that cause the

pumps to stop or slow down when a pressure sensor on the column switching valve

detects an overpressure event.

Flow Paths of the Column Switching Valve

During a manual run, you can change the flow path of the column switching valve by

accessing its dialog box in ChromLab.

To change the flow path of the column switching valve

1.

Touch or click the Column Switching Valve module in the fluidic scheme to
access its dialog box.

2.

In the Column Switching Valve dialog box, select the port to which your column
is connected or through which you want to direct the flow of buffer.