pH Probe

Instrument Guide

| 217

pH Probe

Storing the pH Electrode

When not in use, store the pH electrode in storage solution. See Table

15 on

page

155 for recommended storage solutions.

Tip:

Set the pH valve to Bypass mode until it is needed. This ensures the pH

electrode remains in storage.

To store the pH electrode

1.

Start ChromLab software.

2.

Choose Tools > Calibrate to open the Calibration dialog box.

3.

Select pH on the Calibrate dropdown list.

4.

Press Start to set the pH monitor to Calibration mode.

5.

On the pH valve, inject approximately 10 ml of pH electrode storage solution
through the Cal In port.

6.

Click Close to close the Calibration dialog box.

Note:

You might see a ChromLab message warning you about

interrupting calibration. You can safely ignore this message. Any previous

calibration performed on the pH valve is saved.

Cleaning the pH Electrode

To remove soluble contaminants

Sequentially immerse the electrode in the following solutions for 1 min each:

0.1 M HCl

Distilled H

2

O

0.1 M NaOH

Distilled H

2

O

0.1 M HCl