Bio-Rad ChromLab™ Software User Manual
Page 244
Advertising

A
| Maintaining the Instrument
242
| NGC Chromatography Systems and ChromLab Software
11. Using the 7/64 hex key provided in the accessories kit, remove the cover from
the back of the touch screen monitor.
12. Using a Phillips screwdriver, remove the six screws that attach the touch screen
to the pivot arm assembly and place them aside.
13. Lift the pivot arm assembly off of the touch screen. The mounting assembly
should be in your right hand.