| Connecting the C-96 Autosampler to NGC Systems

294

| NGC Chromatography Systems and ChromLab Software

7.

In the fluidic scheme, touch or click the System Pump module to open its
dialog box.

8.

In the System Pump dialog box:

a.

Set Flow Rate to 5 ml/min.

b.

(Optional) Set Duration to at least 2 min.

c.

Click Start.