Sensors, Air sensors – Bio-Rad ChromLab™ Software User Manual

Page 82

Advertising
background image

2

| The NGC Instrument

80

| NGC Chromatography Systems and ChromLab Software

Sensors

Air Sensors

The NGC chromatography systems can be fitted with an external air sensor module

that can hold up to four air sensors. This module can be daisy chained to an

extension module that can hold an additional four air sensors, providing a total of

eight air sensors. The air sensors detect air, indicating the end of sample or buffer,

and trigger appropriate events through ChromLab software. The air sensors are

available in two sizes that differ in internal diameter to accommodate the two

different tubing sizes:

The large air sensors are used with the larger diameter 1/8" FEP tubing for

the inlet ports of the pumps and inlet valves.

The smaller air sensors are used with the smaller diameter 1/16" PEEK

tubing.

The systems detect the number of air sensors that are attached to the instrument.

Air sensors can be set to detect air in lines filled with buffer. If air is detected in these

lines, the system pumps stop and the run pauses. You can continue the run after

replenishing the buffer and purging the lines.

Air sensors can also be set to detect air in lines filled with sample. If air is detected

in these lines, the sample pump stops so that it cannot pump air onto the column. If

an air-bubble sensing event occurs during a method run, the run continues to the

next step.