Bio-Rad ChromLab™ Software User Manual

Page 32

Advertising
background image

2

| The NGC Instrument

30

| NGC Chromatography Systems and ChromLab Software

Detailed Attributes

The NGC systems include two system pumps: pump A (the left pump) and pump B

(the right pump). The system pumps are available in two flow rates: 10 ml/min (the

F10 pumps) and 100 ml/min (the F100 pumps).

F10 Pump

Flow rate delivery: 0.001–10 ml/min specified

Operating pressure range: 0–3,650 psi (0–25.2 MPa)

Operating viscosity range: 0.35–10 cP

F100 Pump

Flow rate delivery: 0.01–100 ml/min specified

Operating pressure range: 0–1,450 psi (0–10 MPa)

Operating viscosity range: 0.35–10 cP

System Pump LEDs

Each system pump module is fitted with LEDs.

Green LEDs indicate the pump is in use.

How the System Pumps Function

Each NGC system ships with either two F10 pumps or two F100 pumps. The

reciprocating pistons in each pump synchronize to deliver continuous flow with low

pulsation. On the NGC Quest system (without a buffer blending valve), the system

pumps operate such that the sum of the flow rates of pump A and pump B is equal

to the maximum delivery rate. For example, if the flow of an F10 pump A equals 10

ml/min, then the rate for the F10 pump B must equal 0

ml/min. If the system is set to

deliver 5 ml/min with a gradient mixture of 50% B, pump A and pump B each run at

2.5 ml/min.