Recommended maintenance schedule – Bio-Rad ChromLab™ Software User Manual

Page 154

Advertising
background image

A

| Maintaining the Instrument

152

| NGC Chromatography Systems and ChromLab Software

Recommended Maintenance Schedule

Table

14 lists recommended maintenance tasks for the NGC instrument.

Table 14.

NGC instrument maintenance schedule

Interval

Component

Action

Daily

pH monitor

Calibrate the pH monitor. Start ChromLab
software and select Tools > Calibrate and
choose pH from the dropdown list.

Store the pH electrode in storage solution
when not in use. See Table

15

for

recommended storage solutions.

Tip:

See

Storing the pH Electrode on

page

217

for information about storing

the electrode.

Fluidic system

Inspect the system for leaks.

Flush with water at the end of the day to
remove sample and buffer salts.

Columns

Perform CIP and store in 20% ethanol.

Tip:

System CIP is a phase in

ChromLab.

See the NGC Chromatography Systems and
ChromLab Software User Guide.

Waste containers

Empty the containers.

Weekly

System pumps

Change the pump head wash solution and
prime the system.

Sample pump

Change the pump head wash solution and
prime the system.