Pressure Sensors

The NGC systems can have up to four integrated pressure sensors:

System pressure sensor — located in the mixer module

Precolumn and postcolumn pressure sensors — located in the column

switching valve

These pressure sensors protect the column and media from overpressure. One

pressure sensor measures the pressure before the column to protect the

column hardware. The other sensor measures the pressure after the column

and calculates the pressure difference over the media bed. If the pressure

differential value (Delta P) exceeds the preset limit, either the run pauses or

another action is applied.

Sample pump pressure sensor — located in the sample pump module

Each pressure sensor has an accuracy of ±2% or 2 psi, whichever is smaller.

Temperature Sensor

The NGC systems have an integrated temperature sensor located in the

conductivity flow cell that measures the run temperature. Sensor data is used for

temperature compensation of conductivity and pH.