Removing the Wash Housing

To remove the wash housing

1.

Complete the steps in the section

Removing the Pump Heads on page

159

.

2.

Disconnect inlet and outlet tubing from both the left and right wash housings.

3.

Observe the black arrow on the side of each wash housing. The arrows indicate
the direction of wash flow.

On the left housing, the arrow points upward.

On the right housing, the arrow points downward.

Tip:

Before removing the wash housings, label them Left and Right to

easily identify them when you reassemble the piston wash housing.

Caution:

Be careful not to break or damage the piston.

4.

Carefully remove the wash housing. Pull it straight out from the piston.

5.

(Optional) Remove the left wash housing inlet and outlet check valves.

Tip:

It is not necessary to remove the inlet and outlet check valves unless

you plan to replace them. See

To replace the wash housing check valves

on page

173

.

6.

Store the wash housing in a safe place.