Air sensor tubing length and bubble size – Bio-Rad ChromLab™ Software User Manual

Page 103

Advertising
background image

Connecting the Air Sensor Module

Instrument Guide

| 101

Air Sensor Tubing Length and Bubble Size

The response of the air sensor module is highly dependent on the bubble size and

flow rate, and requires a minimum tubing length between the sensor outlet and the

pump inlet. Table

12 lists recommended tubing lengths for various flow rates and

bubble sizes.

Table 12.

Recommended air sensor tubing lengths

Flow Rate

Bubble Size

1/8" ID FEP Tubing Length

(air sensor to pump inlet)

1 ml/min

15–25 μl

≥6"

5 ml/min

50–100 μl

≥6"

10 ml/min

200 μl

≥6"

100 ml/min

1.05–1.4 ml

≥42"