Washing the Columns

Cleaning a column requires access to ChromLab. You can perform this task from

the ChromLab computer or the NGC touch screen. This procedure explains how to

manually wash columns through the System Control tab in ChromLab. Alternatively,

you can wash columns in the following phases in the Method Editor: System

Preparation, Column Activation, or Column Wash. See the

NGC Chromatography

Systems and ChromLab Software User Guide for more information about these

phases.

Caution:

If you wash the column in the reverse flow direction, set the flow rate

to half the column’s recommended flow rate (keep the pressure below the

recommended maximum).

Important:

Verify compatibility before using a wash solvent. Many wash

solvents are a higher viscosity than the typical eluting solvents. You might need

to reduce the flow rate to avoid exceeding the maximum allowable pressure.

Refer to the column specification sheet for recommended cleaning solutions,

flow rates, column volume rates, and storage solutions.

To wash a column

1.

Immerse the system pump inlet line in the buffer solution.

2.

In the fluidic scheme, touch or click the System Pump module to open its
dialog box.

3.

In the System Pump dialog box:

a.

For the NGC Scout and NGC Discover systems, on the Mode dropdown
list select one of the following:

Gradient via Inlet Valves

Gradient via Blending Valve (High Flow)

b.

Set Flow Rate to 1 ml/min (or manufacturer’s specification).

c.

Set Mode to Isocratic.

d.

Set Duration to 3–5 min (or manufacturer’s specification).