The detectors, Replacing the uv flow cell – Bio-Rad ChromLab™ Software User Manual

Page 198

Advertising
background image

A

| Maintaining the Instrument

196

| NGC Chromatography Systems and ChromLab Software

The Detectors

Replacing the UV Flow Cell

The NGC systems ship with a 5 mm UV flow cell installed. Additional flow cells for

analytical (10 mm) and preparative (2 mm) applications are available. This section

explains how to change or replace the UV flow cell. The procedure is the same for

both the single-wavelength UV and multi-wavelength UV/Vis detectors.

To replace the UV flow cell

1.

On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.

2.

Disconnect the inlet and outlet tubings from the UV cell.

3.

With the screwdriver supplied in the fittings kit, loosen the two bolts on the UV
flow cell.

Single-Wavelength UV Flow Cell