Sample injection loops, Columns, Sample injection loops columns – Bio-Rad ChromLab™ Software User Manual
Page 79
Advertising

Tubing, Loops, Columns, and Fittings
Instrument Guide
| 77
Sample Injection Loops
Columns
The NGC chromatography systems support columns for various chromatography
techniques. ChromLab includes a library of the most common commercial columns
in the Method Editor.
Table 6.
NGC system injection loop details
Injection Loop Size
Scope of Use
Maximum Pressure
Rating
100 μl
Small sample sizes
4,000 psi
250 μl
Small sample sizes
4,000 psi
500 μl
Small sample sizes
4,000 psi
1 ml
Large sample sizes
3,000 psi
2 ml
Large sample sizes
3,000 psi
5 ml
Large sample sizes
3,000 psi
DynaLoop™ 25:
25 ml
Very large sample sizes
1,000 psi
DynaLoop 90:
90 ml
Very large sample sizes
1,000 psi