Conductivity detector, And 10 mm; see table, 69 for specific information – Bio-Rad ChromLab™ Software User Manual

Page 71

Advertising
background image

Detectors

Instrument Guide

| 69

Monitors up to four wavelengths in the UV/Vis range of 190–800 nm

Tip:

Use ChromLab to change the absorbance wavelength values. For

more information, see the

NGC Chromatography Systems and ChromLab

Software User Guide.

UV absorbance range: 0–3,000 mAU

UV linearity: ±5%

Three interchangeable flow cells with path lengths of 2 mm, 5 mm (default),

and 10 mm; see Table

4 for specific information

For information about changing the flow cells, see

Replacing the UV Flow Cell

on page

196

.

Note:

Use the preparative 2 mm flow cell to measure high concentrations

of protein. The path lengths of the analytical 5

mm and 10 mm flow cells

are longer and are appropriate for measuring lower concentrations of

protein.

Conductivity Detector

Detailed Attributes

Factory specified range: 0.01–999.9 mS/cm

Validated working range: 0.01–300 mS/cm

Accuracy: ±2%

Flow cell volume: 6

μ

l

Integrated temperature sensor range: 4–100

°C

Table 4.

Pressure limits of UV detector flow cells

Pump Size

Pressure

2 mm

5 mm

10 mm

F10

500 psi

F100

500 psi