Site requirements, Power considerations – Bio-Rad ChromLab™ Software User Manual

Page 16

Advertising
background image

1

| Introduction

14

| NGC Chromatography Systems and ChromLab Software

Extend the preconfigured systems with additional valves for buffers,

samples, and columns

Organize the location of the modules to optimize separation performance

based on method scale and complexity, and to minimize the system swept

volume

Minimize errors when connecting tubing using the Point-to-Plumb™

feature in ChromLab software

Site Requirements

Power Considerations

Note:

The power supply to the NGC system must be stable and within

specifications at all times to ensure optimal operation. The power cable

connected to the power inlet port must be rated for 7 amps or more.

Table 1.

Power requirements for the NGC system and ChromLab

Parameter

Requirement

Mains input voltage

AC 100—240 V; 50—60 Hz

NGC maximum power usage

<750 watts

Number of power sockets

Minimum of three power sockets:

One socket for the NGC instrument

One socket for the computer running
ChromLab

One socket for the BioFrac™ fraction
collector

A socket for any supported peripheral
instrument that you attach (for example,
the C-96 autosampler)

Tip:

The touch screen does not

require a separate power supply.

Type of power sockets

IEC type connections, grounded sockets