Plumbing and Priming the C-96 Autosampler

This section explains how to plumb the C-96 autosampler to the sample inject valve

on the NGC instrument and to prime the autosampler after you plumb it. Before you

plumb and prime, you must prepare the tubing. Use 1/16" PEEK tubing to plumb the

autosampler to the sample inject valve.

For information about plumbing the autosampler, see the C-96 Autosampler for

BioLogic DuoFlow™ Chromatography System Quick Start Guide or the C-96

Autosampler for BioLogic DuoFlow Chromatography System User Manual.

To prepare the tubing

1.

Cut two segments of 1/16" PEEK tubing of adequate length.

2.

To one end of each segment, attach a 10-32 fitting. This fitting attaches to the
autosampler.

3.

To the other end of each segment, attach a 1/4-28 fitting. This fitting attaches to
the sample inject valve.

To connect the tubing from the autosampler to the sample inject valve

1.

Ensure that both the NGC instrument and the C-96 autosampler are powered
off.

2.

Attach one 10-32 fitting to Port 1 and the other 10-32 fitting to Port 6 on the
autosampler.

3.

Attach the 1/4-28 fitting from Port 1 to the Loop E port on the sample inject
valve.

4.

Attach the 1/4-28 fitting from Port 6 to the Loop F port on the sample inject
valve.

5.

Insert one 1/4-28 plug to the syringe port and another 1/4-28 plug to the Waste
port (W1) on the sample inject valve.

6.

Power on the NGC instrument and the autosampler.