3

| Preparing the Instrument

104

| NGC Chromatography Systems and ChromLab Software

Connecting External Devices to the NGC Instrument

Tip:

If you do not plan to connect the C-96 autosampler or other external

devices to the NGC instrument you can skip this section.

This section explains how to connect the signal import module (SIM) to the NGC

instrument, and how to connect external devices to the SIM.

For information about connecting the C-96 autosampler to the SIM, see

Appendix D, Connecting the C-96 Autosampler to NGC Systems

on

page

289

.

For information about connecting the SIM to an NGC instrument, see

Connecting the SIM to the NGC Instrument on page

105

.

For information about connecting an external device to the SIM, see

Connecting External Devices to the SIM on page

107

. Next, see the

chapter System Controls in the NGC Chromatography Systems and

ChromLab Software User Guide to enable the external device.

For information about installing and setting up an external device, see its

installation documentation.

Controlling Programmed Functions via
Transistor-to-Transistor Logic

You can control a function of an external device or receive function information from

a device via the SIM. Using the SIM’s transistor-to-transistor logic (TTL), the NGC

instrument can mark when a function occurred on an external device through the

SIM’s DIGITAL IN port, or send commands to an external device through its DIGITAL

OUT port.

For example, you can connect a C-96 autosampler to the SIM enabling NGC

systems to provide automated, accurate, and reproducible sample injections.

Connecting the autosampler to the SIM’s DIGITAL OUT and DIGITAL IN ports

enables the NGC instrument to inform the autosampler when to begin its injection

cycle and to mark when the autosampler injects a sample.