Calibrating the pump flow rate, Calibrating pressure sensor zero point – Bio-Rad ChromLab™ Software User Manual

Page 148

Advertising
background image

3

| Preparing the Instrument

146

| NGC Chromatography Systems and ChromLab Software

Calibrating the Pump Flow Rate

Prime and purge the system pumps with water before starting the calibration

procedure. Calibrate the flow rate with water.

Note:

Ensure that the column is offline before performing this task.

To calibrate the pump flow rate

1.

Do one of the following:

On the ChromLab computer, select the System Control tab, click

Tools

>

Calibrate, and select Pump Flow Rate on the dropdown list.

On the NGC touch screen, select Calibrate on the ChromLab dropdown

menu and then select Pump Flow Rate on the dropdown list.

2.

Follow the steps in the Pump Flow Rate Calibration dialog box.

3.

Verify that Status (at the bottom of the dialog box) is Calibrated.

Calibrating Pressure Sensor Zero Point

Use this procedure to zero the pressure on the following modules:

Sample pump

System pump

Column switching valve

To zero the pressure

1.

Ensure the modules are not plumbed and they are at ambient pressure.

2.

Do one of the following:

On the ChromLab computer, select the System Control tab, click

Tools

>

Calibrate, and select the pump or valve on the dropdown list.

On the NGC touch screen, select Calibrate on the ChromLab dropdown

menu and then select the pump or valve on the dropdown list.